XDC-2 Typical Operating Specifications (1.)

  • Temperature Range: 1,075K to 1,925K (800°C to 1,650°C)
  • Temperature Resolution: + 0.5% at over 800°C, Type B Thermocouples
  • Configuration: Horizontal Differential Measurement Module
  • Thermal Expansion:
    Sample length: Up to 50 mm
    Sample diameter: 10 mm to 19 mm
    Sample support: Alumina and Sapphire material
    LVDT range: + 6.35mm
    LVDT resolution: + 0.1 μm
    LVDT range accuracy: + 1.9 μm (2.)
    Heating/cooling rate 150°C/hour or better

  • Thermal Diffusivity:
    Range: 0.01 cm2/s to 1 cm2/s
    Resolution: +0.01 cm2/s
  • Thermal Conductivity:
    Range: 0.01 W/m K to 500W/m K
    Resolution: +10% or better
  • Test Environment:
    Oxidizing or inert
    Air, N2, CO2, He, Ar, Vacuum purge (3.)
    Digital Thermal Mass Flow Meter/Controller, 3 units
  • Data Interface:
    Network: RJ45, 10-100baseT LAN
  • Power and Cooling:
    208VAC, single phase, 45A service, other voltages available upon request
    Water, up to 5 GPM at 10-37°C
  • Size and Weight: (LWH, pounds)
    Measurement System: 72” x 30” x 58”, approximate weight 900#
    Minimum operating clearance required: 96” x 36” x 96” Shipping: 80” x 42” x 55”, Approximate shipping weight 1,200#

Notes:

  1. Specifications subject to change without notice
  2. Accuracy is improved over smaller temperature ranges